User Contributed MET/CAL PROCEDURE ============================================================================= INSTRUMENT: SD104A-1 Sweep Rate Sub DATE: 08-Sep-95 AUTHOR: User Contributed REVISION: ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 6 NUMBER OF LINES: 79 CONFIGURATION: HP 5335A ============================================================================= STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 HEAD {} 1.002 HEAD {ÉÍÍÍÍÍÍÍÍÍÍÍÍÍ»} 1.003 HEAD {º Sweep Times º} 1.004 HEAD {ÈÍÍÍÍÍÍÍÍÍÍÍÍͼ} 1.005 HEAD {Linear Sweep, Rate=10} 1.006 HEAD ÍÍ Sweep Times ÍÍ 1.007 JMP 2.001 1.008 EVAL 2.001 DISP ÿÿÿÿÿÿÿÿSet the UUT as follows: 2.001 DISP Mode Switch ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ[16] SINGLE LINEAR 2.001 DISP SWEEP RATE ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ[16] 10 2.001 DISP MULTIPLIER (B) ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ[16] 10 2.002 DISP Connect the J23 on the UUT to CH A of the 2.002 DISP HP 5335A counter. 2.003 IEEE [@5335][TERM LF] 2.004 IEEE [@5335]IN[D1000]; 2.005 IEEE [@5335]FN2;GM0;GA.25;CO1; 2.005 IEEE [@5335]TR1;AT3.5;BT2; 2.006 DISP On the UUT depress the RESET DOWN button. The LED above 2.006 DISP the button should be on. Now depress the SWEEP UP 2.006 DISP button. That LED should go on and the RESET DOWN LED 2.006 DISP should extinguish. 2.007 HEAD [27][91]5m**PLEASE WAIT. LONG GATE TIME!**[27][91]m 2.008 IEEE [@5335][T0][I] 2.009 HEAD 2.010 ACC 49.95T 0.001% 2.011 MEME 2.012 MEMC T 0.500U mult=10 #! Test Tol 0.5, Sys Tol 0.0004995, TUR 1001.001 (>= 4.00). 3.001 HEAD ÍÍ Sweep Rate ÍÍ 3.002 DISP ÿÿÿÿÿÿÿÿSet the UUT as follows: 3.002 DISP MULTIPLIER (B) ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ[16] 1 3.003 IEEE [@5335]IN[D1000]; 3.004 IEEE [@5335]FN2;GM0;GA.25;CO1; 3.005 IEEE [@5335]TR1;AT3.5;BT2; 3.006 DISP On the UUT depress the RESET DOWN button. The LED above 3.006 DISP the button should be on. Now depress the SWEEP UP 3.006 DISP button. That LED should go on and the RESET DOWN LED 3.006 DISP should extinguish. 3.007 HEAD [27][91]5m**PLEASE WAIT. LONG GATE TIME!**[27][91]m 3.008 IEEE [@5335][T0][I] 3.009 HEAD 3.010 ACC 499.5T 0.001% 3.011 MEME 3.012 MEMC T 5.00U mult=1 #! Test Tol 5, Sys Tol 0.004995, TUR 1001.001 (>= 4.00). 4.001 HEAD {} 4.002 HEAD {Log Sweep, Mult.=10, Rate=10} 4.003 HEAD ÍÍ Sweep Rate ÍÍ 4.004 JMP 5.001 4.005 EVAL 5.001 DISP ÿÿÿÿÿÿÿÿSet the UUT as follows: 5.001 DISP Mode Switch ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ[16] SINGLE LOG 5.001 DISP MULTIPLIER (B) ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ[16] 10 5.002 IEEE [@5335][TERM LF] 5.003 IEEE [@5335]IN[D1000]; 5.004 IEEE [@5335]FN2;GM0;GA.25;CO1; 5.005 IEEE [@5335]TR1;AT3.5;BT2; 5.006 DISP On the UUT depress the RESET DOWN button. The LED above 5.006 DISP the button should be on. Now depress the SWEEP UP 5.006 DISP button. That LED should go on and the RESET DOWN LED 5.006 DISP should extinguish. 5.007 HEAD [27][91]5m**PLEASE WAIT. LONG GATE TIME!**[27][91]m 5.008 IEEE [@5335][T0][I] 5.009 HEAD 5.010 ACC 18.0T 0.001% 5.011 MEME 5.012 MEMC T 0.10U #! Test Tol 0.1, Sys Tol 0.00018, TUR 555.556 (>= 4.00). 6.001 END #! T.U.R.s less than 4.00: 0 #! T.U.R.s estimated using RANGE value: 0 #! T.U.R.s not calculated (ASK- U): 0 #! T.U.R.s not computable at compile time: 0 #! FOR JUSTIFICATION REFER TO COMMENTS FOLLOWING EACH TEST IN THIS LISTING.